Инструкция терминала 715(Avery Weigh-Tronix)
FIGURE 1
Hidden E Key on Model 715 Indicator
Viewing Current Configuration Code
Entering a New Configuration Code
You may exit this procedure from any display, except E, with GROSS, HOLD, or LOAD/ UNLOAD key.
Step 1 display depends upon selected capacity for your system.
l. | Key in 715. . . | 715 is displayed. |
2. | Press hidden E key and hold for two seconds. . . |
715E is displayed. |
3. | Press E key. . . | E is displayed. |
4. | Press LOAD/UNLOAD. . . | Current Configuration is displayed. |
5. | Press GROSS, HOLD, or LOAD/UNLOAD key to return to corresponding operational mode. |
Step 2 display depends upon selected capacity for your system.
Determine an appropriate Configuration Code number from chart in TABLE l.
Key in 7 1 5. . . 715 is displayed.
Press hidden E key and hold for
two seconds. . . 715E is displayed.
Key in appropriate 4- or 5-digit
Configuration Code. . . New Configuration Code is dis-
played.
(See Table 2: Valid Numbers for Configuration Code.) If re-keying is necessary, press ZERO/RESET, then re-key.
Press E key. . . E is displayed.
Press LOAD/UNLOAD. . . New Configuration Code is displayed
again; entry of code is complete.
Press GROSS, HOLD, or LOAD/UNLOAD key to return to corresponding operational mode. (Refer to TABLE 1 on following page.)
Axle Sizes
Using Table 1: Configuration Code Example
Physical axle size and calibration axle size are not the same for some Weigh Bars. If your Weigh Bar physically measures 5/8" in diameter but is marked with a calibration axle size of 2-1/4" D (Dual), then use data shown in Table 1 for a 5/8" bar. If your Weigh Bar measures 1" in diameter but is marked with a calibration axle size of 2-1/8 ", use data shown in Table 1 for a 1" bar.
For Weigh Bars or weight transducers which have a physical axle size of either 5/8" or 1" but which have a calibration axle size other than 2-1/4" D or 2-1/8", respectively, select Custom Sensitivity and Capacity x Increment in Table 1.
For all other Weigh Bars or weight transducers, if physical axle size and calibration axle size of a bar do not match, use the calibration axle size for Table 1.
Table 1 shows you how selections shown for these parameters would be represented by a Configuration Code of 5151:
CONFIGURATION CODE EXAMPLE | ||
Parameter | Represented by | Selection |
Axle Size | 2nd digit | 2-1/8" |
Capacity X Increment Size | 3rd digit | 20K x 2 |
Warning Alarm Factor | 4th digit | 25 x 2 = 50lb |
Units of Measure | 5th digit | lb |
Auto Hold Mode | 5th digit | ON |
CODE 5151 uses 2nd, 3rd, 4th and 5th of five available digits. |
Divide desired WARNING ALARM VALUE by Increment Size to find
WARNING ALARM FACTOR.
Table 1: Configuration Code Chart
CALIBRATION AXLE SIZE+ |
CAPACITY x |
INCREMENT SIZE | ||||
0: 5/8"* | 200 x 0.01 | 200 x 0.02 | 200 x 0.05 | 2K x 0.1 | 2K x 0.2 | 2K x 0.5 |
1: 1"* | 2K x 0.1 | 2K x 0.2 | 2K x 0.5 | 20K x 1 | 20K x 2 | 20K x 5 |
2: 1-1/4" | 2K x 0.1 | 2K x 0.2 | 2K x 0.5 | 20K x 1 | 20K x 2 | 20K x 5 |
3: 1-7/8" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
4: 2" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
5: 2-1/8" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
6: 2-1/4" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
7: 2-1/4"D | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
8: 2-1/4"D-P | 200K x 10 | 200K x 20 | 200K x 50 | 200K x 100 | 200K x 200 | 200K x 500 |
9: 2-1/2" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
10: 3-1/8" | 200K x 10 | 200K x 20 | 200K x 50 | 200K x 100 | 200K x 200 | 200K x 500 |
11: 4" | 200K x 10 | 200K x 20 | 200K x 50 | 200K x 100 | 200K x 200 | 200K x 500 |
12: CC-20 | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
13: Alley Weigh | 2K x 0.1 | 2K x 0.2 | 2K x 0.5 | 20K x 1 | 20K x 2 | 20K x 5 |
19: Custom Sensitivity and Capacity x Increment (See text: Configuring Custom Sensitivity with Capacity x Increment Size.)
Numbers 0 1 2 3 4 5
0: 0
1: 2
2: 5
3: 10
4: 20
5: 25
6: 30
7: 40
8: 50
9:Custom (See text: Configuring Custom Warning Alarm Factor.)
0: lb No
1: lb Yes
2: kg No
3: kg Yes
_ 5 1 5 1 is the Configuration Code number representing selections cited in USING TABLE 1: CONFIGURATION CODE EXAMPLE.
1 2 3 4 5 are place numbers for each digit in Configuration Code.
+ Axle size determines Weigh Bar sensitivity in mV/V. Refer to Table 3 to relate axle sizes to exact sensitivity values.
K = Multiply by 1000
* If physical axle size of your Weigh Bar equals 5/8" or 1", but differs from its identifying calibration axle size, see text: AXLE SIZES.
D-P =Dual Platform: Eight 350-ohm Weigh Bars and junction box with span pots.
D =Dual: double-ended Weigh Bar
@ Divide desired Warning Alarm by selected Increment Size to find Warning Alarm Factor.
Table 2 Valid Numbers for the Configuration Code | ||
PARAMETERS VALID #'S INVALID #'S | ||
Axle Size |
0 - 13, 19 |
14 - 18 |
Capacity x Increment Size |
0 - 5 |
Any number greater than 5 will be recognized as a 5. |
Warning Alarm Factor |
0 - 9 | |
Units and Auto Hold |
0 - 3 |
Any number greater than 3 will be recognized as a 3. |
Table 3 Model 715 Calibration Table | |||||
Calibration Axle Size |
Scale Capacity |
Number of Weigh Bars |
mV/V |
Calibration | |
5/8"* | 200 lb | 4 | 0.2 | 180.039 | lb |
1"* | 2,000 lb | 4 | 0.4 | 1,464.82 | lb |
1-1/4" | 2,000 lb | 4 | 0.2 | 1,742.58 | lb |
Alley Weigh | 2,000 lb | 2 | 1.0 | 1,913.8 | lb |
1-7/8" | 20,000 lb | 4 | 0.6 | 15,399.2 | lb |
2" | 20,000 lb | 4 | 0.6 | 19,143.8 | lb |
2-1/8" | 20,000 lb | 4 | 0.4 | 14,648.2 | lb |
2-1/4" | 20,000 lb | 4 | 0.4 | 18,003.9 | lb |
2-1/4"D | 20,000 lb | 4 | 0.2 | 18,003.9 | lb |
2-1/2" | 20,000 lb | 4 | 0.2 | 12,540.9 | lb |
2-1/4"D | 200,000 lb | 4 | 1.0 | 90,019.3 | lb |
2-1/4"D-P | 200,000 lb | 8/350 | 1.0 | 186,816.0 | lb |
3-1/8" | 200,000 lb | 4 | 1.0 | 126,402.0 | lb |
4" | 200,000 lb | 4 | 0.6 | 155,273.1 | lb |
CC-20 | 200,000 lb | 4 | 2.0 | 109,940.0 | lb |
* If physical axle size of your Weigh Bar equals 5/8" or 1", but differs from its identifying calibration axle size, see previous text: AXLE SIZES. D-P is Dual Platform: Eight 350-ohm Weigh Bars and junction box with span pots. D is Dual: double-ended Weigh Bar. |
Configuring Custom Sensitivity with Capacity x Increment Size
You may exit this procedure from any display, except E, with GROSS, HOLD, or LOAD/ UNLOAD key.
This procedure allows you to select a weight transducer sensitivity rating other than those referenced by axle sizes in Table 1. You may even config- ure another manufacturer's weight transducer for interface with the Model 715. Before starting the keying procedure for custom configuration, you must verify three facts about your interfaced transducer(s):
Trade name of transducer
Sensitivity of transducer
Capacity x Increment Size of transducer
TRADE NAME OF TRANSDUCER
To configure custom sensitivity for Weigh-Tronix Weigh Bars, continue now to the next topic, SENSITIVITY.
To configure for transducers from manufacturers other than Weigh-Tronix, refer now to instructions located in the PROGRAMMING APPENDIX. Then, if you find that custom configuration of sensitivity is required, return to this section and proceed to the next topic.
SENSITIVITY OF TRANSDUCER
Determine the sensitivity rating, that is, mV/V output at full capacity, for the weight transducer(s) you are using. The transducer probably carries this information, or access to it, stamped or printed on the transducer itself.
Call the factory at 1-800-458-7062 for help in determining the Model 715 sensitivity rating for your transducer(s).
Because many different types of bars are used by various manufactur- ers, it is impossible to generically document a way to determine Model 715 sensitivity ratings for all available weight transducers.
Write down your calculated sensitivity rating: , using
this format:
n.nnnnnnn = your sensitivity rating 1 2345678 = digit places
You may key in a mV/V value of up to eight digits. Selection range for sensitivity is from 0.075 to 6.0 mV/V. Keep your notation handy for entry in steps 6 and 8 of the following keying procedure.
CAPACITY X INCREMENT SIZE OF TRANSDUCER
Refer to TABLE 4 and locate a Capacity x Increment Size combination that is appropriate for your Indicator.
Note the column number in which appropriate Capacity x Increment Size combination is located. Write down the column number: Keep
your notation handy for entry in step 3 of the following keying procedure.
Table 4 Row Numbers for Custom Capacity x Increment Size | ||||||
Column Numbers: 0 |
1 |
2 |
3 |
4 |
5 | Row #’s |
200K x 10.0 |
200K x 20.0 |
200K x 50.0 |
200K x 100.0 |
200K x 200.0 |
200K x 500.0 |
0 |
20K x 1.0 | 20K x 2.0 | 20K x 5.0 | ||||
2K x .1 | 2K x .2 | 2K x .5 | ||||
200 x .01 | 200 x .02 | 200 x .05 |
Now look to the right, and find the row number that applies to your selected Capacity x Increment Size.
Write down the appropriate row number for use in step 12: . Steps 1-5 open the door to custom configuration of Sensitivity and Capacity x Increment Size.
Key in 715. . . 715 is displayed.
Press hidden E key and hold for
two seconds. . . 715E is displayed.
Key in custom Configuration Code:
For digits 1 and 2, key in 19 for Custom (See TABLE 1).
For digit 3, key in column number you previously identified in Letter E.
For digits 4 and 5, key in appropriate code numbers for Warning Alarm Factor, Units of Measure, and Auto Hold parameters.
New Configuration Code is dis- played.
Press E key. . . E is displayed.
Press LOAD/UNLOAD. . . New Configuration Code is displayed
again.
First four digits of custom sensitivity are configured into Register 1 by steps 6, 7, and 8.
Key in first four digits of sensitivity rating, previously identified in Letter C,
n.nnnnnnn 1 234
pressing 0 key for each unused
digit place. . . First four digits are displayed.
Press E key. . . E is displayed.
Press 1 key. . . First four digits are again displayed.
Last four digits of custom sensitivity are configured into Register 2 by steps 9, 10, and 11.
Configuring Custom Warning Alarm Factor
You may exit this procedure from any display, except E, with GROSS, HOLD, or LOAD/UNLOAD key.
Key in last four digits of appropriate sensitivity rating,
n. nnnnnnn 5678
pressing 0 key for each unused digit place or once for four unused digit places. . . Entered digits are displayed.
Press E key. . E is displayed.
Press 2 key. . . Entered digits, last four digits of sensitivity rating, are again dis- played.
Key in row number you previously
identified in Letter G. . . Row number is displayed.
Press E key. . . E is displayed.
Press 3 key. . . Row number is again displayed.
Configuration of Custom Sensitivity with Capacity x Increment Size is complete.
Press GROSS, HOLD, or LOAD/UNLOAD key to return to corresponding operational mode.
This procedure starts with example Configuration Code 5151 shown in TABLE 1. Code 5151 used the second, third, fourth, and fifth of five avail- able digit places; the first digit place is unused.
We will configure a Custom Warning Alarm Factor of 60 to provide an actual Warning Alarm Value of 120. We divide the desired Warning Alarm Value by the selected Increment Size to find the Warning Alarm Factor. TABLE 1 shows that in Configuration Code 5151, the second digit place, a 5, (fourth digit place from right) selects an Increment Size of 2.
Key in 7 1 5. . . 715 is displayed.
Press hidden E key and hold for two seconds. . . 715E is displayed.
Using example Configuration Code 5151, change fourth digit place (second from right) to the numeral 9, for Custom, and key
in Configuration Code 5 1 9 1 . . . 5191 is displayed.
Press E key . . . E is displayed.
For Step 6, any keyed in value higher than 2000 will be read by the Indicator as 2000.
Selection Range for Custom Warning Alarm Factor is O to 2000.
Press LOAD/UNLOAD . . . 5191 is again displayed.
Key in Custom Warning Alarm
Factor, 6 0. . . 60 is displayed.
Press E key . . . E is displayed.
Press 4 key . . . 60 is again displayed.
Configuration of Custom Warning Alarm Factor is complete.
Press GROSS, HOLD, or LOAD/UNLOAD key to return to corresponding operational mode.
Calibration Mode
Instructions for Calibration Mode include:
Calibrating Using a Known Weight
Getting Ready to Calibrate with BLH Model 625 Calibrator
Calibrating with BLH Model 625 Calibrator
Use a reliable known weight equal to or greater than the maximum weight of
Calibrating Using a Known Weight
You may exit this procedure from any display, except E, with GROSS, HOLD, or LOAD/UNLOAD key.
material that will be weighed.
l. | Empty and zero the scale. | |
2. | Key in 7 1 5. . . | 715 is displayed. Step 2 display depends upon selected capacity for your system. |
3. | Press hidden E key and hold for two seconds . . . |
715E is displayed. |
4. | Press E key. . . | E is displayed. |
. 5. |
Press GROSS . . . |
Zero value should be displayed |
before you continue. If RE-ZERO- ING is necessary following Step 5, press ZERO/RESET. | ||
6. | Place known weight on scale . . . | Display represents gross weighment of known weight prior to calibration. |
7. | Adjust span, using numeric keys as shown, until weight display equals value of known weight (Span is adjustable +/- 10%): |
To increase span, press 1, 2, 4, or 5;
To decrease span, press 6, 7, 9, or 0;
To return span to factory-calibrated value, press 3 or 8. . .
Calibration is complete.
Press GROSS, HOLD, or
LOAD/UNLOAD key to return
to corresponding operational mode.
Getting Ready to Calibrate with BLH Model 625 Calibrator
This method of calibration is intended for trained technicians using special- ized equipment.
Use appropriate instruction shown in TABLE 5 to connect the BLH Model 625 Calibrator to the Model 715 Indicator.
Determine highest mV/V reading according to factors of Weigh Bar size and full capacity of system, using appropriate Calibration Table as referenced in TABLE 5.
Write down highest mV/V reading for use in Step 7 of following proce- dure: .
Table 5 Calibrator Connections And Calibration Tables | ||
Number of Weigh Bars |
Instructions for Connecting | |
and Type of Connector | Indicator with BLH Calibrator | Calibration Tables |
Four Weigh Bars w/ | Connect single cable from BLH | Use W-T Spec P/N 24375 |
5-Pin Connectors | Calibrator to one of the | Model 715 Calibration |
5-pin Weigh Bar connectors. | Tables. | |
Four Weigh Bars w/ | Connect four cables from a | Use W-T Spec P/N 14599 Farm |
4-Pin Connectors | modified BLH Calibrator to all | Scale Calibration Tables. |
four 4-pin Weigh Bar connectors. | ||
One Weigh Bar w/ | Connect single cable from BLH | Use W-T Spec P/N 24375 |
7-pin Connector | Calibrator to the single 7-pin | Model 715 Calibration |
Weigh Bar Connector. | Tables. |
Calibrating with a BLH Model 625 Calibrator
You may exit this procedure from any display, except E, with GROSS, HOLD, or LOAD/UNLOAD key.
Dial BLH Calibrator to zero mV/V.
Zero Indicator.
Key in 7 1 5 . . . 715 is displayed. Step 3 display depends upon selected capacity for your system.
Press hidden E key and hold for two seconds . . . 715E is displayed.
Press E . . . E is displayed.
Press GROSS. . . Zero value should be displayed before you continue. IF RE-ZERO- ING is necessary following Step 6, press ZERO/RESET.
Turn BLH Calibrator dial to highest mV/V setting for this system, as determined in Step 2 of Getting Ready to Calibrate
with BLH . . . Display represents gross weight prior to calibration.
Adjust span, using numeric keys as shown, until weight display equals weight value known in Calibration Table for highest mV/V setting (Span is adjustable to +/- 10%):
Press 1, 2, 4, or 5 to increase span;
Press 6, 7, 9, or 0 to decrease span;
Press 3 or 8 to return span to factory calibrated value. . .
Calibration is completed. HOLDING IN NUMERIC KEY in Step 8 causes display progression to auto-repeat and accelerate.
Press GROSS, HOLD, or LOAD/UNLOAD key to return to corresponding operational mode.
Auto Zero Tracking Selection
Disabling Azt
Enabling Azt
Model 715 comes standard with Auto Zero Tracking (AZT) of +/- .025% of full capacity. This section explains how to disable this feature if necessary.
Key in 715 . . . 715 is displayed.
Press hidden E key and hold for
two seconds . . . 715E is displayed.
Press 0 . . . 0 is displayed.
Press the hidden E key . . . E is displayed.
Press and hold 0 until . . . 0 is displayed.
Key in 715 . . . 715 is displayed.
Press hidden E key and hold for two seconds . . . 715E is displayed.
Press 1 . . . 1 is displayed.
Press hidden E key . . . E is displayed.
Press and hold 0 until . . . 1 is displayed.
Internal Raw Counts Mode
Use this mode to access Internal Raw Counts display. Internal Raw Counts value is useful in two ways:
In troubleshooting, to verify correct A-D output.
In conjunction with a coarse zero potentiometer on a junction box, to adjust absolute zero.
Viewing Internal Raw Counts Display
You may exit this procedure from any display, except E, with GROSS, HOLD, or LOAD/UNLOAD key.
Software Revision Mode
l. Key in 715. . . 715 is displayed.
Press hidden E key and hold for two seconds . . . 715E is displayed.
Press E. . . E is displayed.
Press ZERO/RESET. . . Internal Raw Count value is dis-played.
At 1.0 mV/V, the display will always read 10,955 counts, +/- 6%.
You May Adjust Absolute Zero if your scale system includes a junction box with a coarse zero potentiometer. When Internal Raw Counts are displayed, remove all weight from the scale and adjust the coarse zero potentiometer until the display reads 0.
Press GROSS, HOLD, or LOAD/UNLOAD key to return to corresponding operational mode.
Use this mode to verify the software version that has been installed in your 715.
Key in 715 . . . 715 is displayed.
Press hidden E key and hold for
two seconds . . . 715E is displayed.
Press hidden E key . . . E is displayed
Press 5 and hold . . . XXX is displayed.
Release key . . . Rev. X is displayed.
This information gives you the dash number of the Weigh-Tronix 27657 - XXXX with appropriate revision level.
Circuit Descriptions
Main Printed Circuit Card
This printed circuit card contains the
Power Supply
Analog-To-Digital Circuitry
Microprocessor Circuitry
Display Circuitry.
To prevent the unit from turning off below 10 volts, remove CR5 and C36. When configured this way, the unit will return to the Gross mode if it loses power.
This indicator, which is energized by 10 to 18 vdc in reference to a NEGA- TIVE GROUND utilizes two five amp fuses mounted on the bottom of the chassis for current protection. (Systems with a positive ground are modified by cutting jumpers from E11 to E12 and E9 to E10. Then install jumpers from E9 to E12 and E10 to E11.)
The circuit card receives power from the fuse at P2 and is protected against high voltage transients with varistor RV1.
Input voltage is monitored by 16 volt zener diode VR2. If the voltage across VR2 exceeds 16 vdc, Q3 will turn “ON” holding the power circuitry in the “OFF” condition. In addition, if the input voltage drops below 10 volts DC, PA7 disables the power circuitry by outputting a low signal.
A “Power On” command is generated when the ON key is pressed. This turns Q5 on and provides a low signal to +8v regulator (u18) pin1 enable input. When the power is turned off, the Q5 turns off, the Q4 turns on and a high signal disables (U18) the +8 regulator and turns it off.
The -7.5 vdc supply voltage is established by U7, a voltage convertor. This device inverts the input voltage to a -7.5 vdc level.
The +5 vdc voltage is provided by U8, a 7805 regulator.
As soon as the +8 volt power supply voltage is present, Q2 turns on which turns on the DS2 and DS3 LEDS
Analog to Digital Circuitry
The analog-to-digital circuitry consists of two weight voltage amplifiers (U1 and U2), a reference amplifier (U3), an analog switch (U4), switch control logic gate (U14C), a dual slope integrator (U5), and a comparator (U6).
An 8 vdc excitation voltage is supplied to the weight sensors from P1 pins 1 and 6. The millivolt signal received back from the weight sensors enters the analog-to-digital circuitry at P1, pins 3 and 4.
This millivolt signal is amplified and filtered by U1 and U2 to establish the amplified weight voltage (0 to +9 vdc) at U2 pin 6. This weight voltage is then provided to U4 pin 5. A logic low during the weight voltage interval at U4 pin 1 enables the amplified weight voltage to be received by the dual slope integrator (U5). Starting at zero volts, the integrator capacitor (C20) charges at ramp angle directly proportional to the amplified weight voltage but of opposite polarity. As the amplified weight voltage increases, the integrator output ramps more in the negative direction.
The voltage received from the sense lines enters the analog-to-digital circuitry at P1 pins 2 and 5. This sense voltage establishes a reference voltage of -9 vdc at U3 pin 6. The reference voltage is applied to U4 pin 10.
After the integrator capacitor has been charged for a fixed period of time, the Weight Voltage Interval, U4 pin 1 goes high. U14 applies a logic low to pin 14 of U4 which allows the reference voltage to be applied to the integra- tor. Since the reference voltage (-9 vdc) is of opposite polarity to the ampli- fied weight voltage, it causes the integrator output to ramp back toward zero. (This period of time is known as the Reference Voltage Interval.)
While the integrator output is ramping back to zero, the comparator (U6) signals the microprocessor to record this period of time in clock pulses. The microprocessor uses this number of pulses to calculate the amount of weight on the scale.
When the integrator reaches zero, the comparator output goes low which triggers U14C to remove the logic low at U4 pin 14 completing one analog- to-digital cycle.
This analog-to-digital conversion repeats at a frequency of 30 hertz.
Microprocessor Circuitry
The Microprocessor Circuitry consists of a Motorola MC68HC11A1 8-bit microprocessor IC (u11), a 74HC573 tristate latch (U12), a 27C128 EPROM (U13), and various other components.
The microprocessor features on chip memory 256 bytes of RAM, 512 bytes of EEPROM, along with a Computer Operating Properly (COP) watch dog system.
A 2 MHZ internal clock is contained in the microprocessor and is controlled by an external crystal (Y1). Output (ECLK) from this clock provides timing reference for the EPROM (U13), and input to voltage monitoring circuit U9.
The microprocessor communicates with Analog-to-Digital switch control logic gate U14 through ports PA2 and PA6. Through U14, the timing refer- ence for the Reference Voltage Interval and Weight Voltage Interval are provided.
The switches on the front panel receive a signal from ports PD2-PD5 on the microprocessor and they are monitored for an electrical change by ports PE0PE3 on the microprocessor.
As signals are received by the microprocessor, it will send them to the EPROM (U13) on the address lines A0-A15. The signals will be translated and held by U12. Once all translating is completed, the microprocessor will enable U12 to return the new information on data lines D0-D7 to the micro- processor. Then, the appropriate device will receive a signal from the microprocessor.
The Computer Operating Properly (COP) watch dog system in the micropro- cessor detects errors introduced by electromagnetic interference. If an error is detected, the microprocessor is reset and operation continues. When the indicator is powered up, the momentary displayed "hi" message, indicates the watch dog system is active.
If a "hi" message is displayed, the indicator will perform correctly, but will not be guarded by the COP watch dog system. This may happen when a new microprocessor IC is installed. The COP watch dog system can be activated by following this procedure:
Turn the indicator OFF jumper P7-1 to P7-2 and P6-1 to P6-2
Turn the Model 715 ON for a few seconds
Turn the Model 715 OFF
Remove the jumpers
Display Circuitry
Optional RS-232 or Remote Circuitry
To Install Remote Receiver
The six digit liquid crystal display (DS1) features black segments against a silver transflective background.
Liquid crystal display driver U15 requires a +5 vdc power supply and two signal inputs from the microprocessor to function: (1) a 33-bit serial data input and (2) a microprocessor controlled input. When both of these condi- tions have been met, the display driver will energize the appropriate seg- ments on the display.
Additional circuitry required to provide either an RS-232 communications interface or a remote display interface requires R49-57, C48-51, U16, U10, and P6 connector.
The RS-232 transmitter/receiver (U10) internally generates +10v and -10 v supplies and meets all EIA RS-232C specifications. Through U10, the microprocessor communicates with peripherals which will accept the follow- ing protocol: baud rate 1200, 1start bit, 8 data bits, no parity, and one stop bit. The RS-232 communications port supports X-on (HEX 11), X-off (HEX 13_) handshake and responds to an inquire character (HEX 05) print request.
The invertor/buffer gates (U16) are used in the remote clock, remote data, remote zero, and remote print lines to provide isolation and assure proper logic; levels to and from the remote display.
P6 connector provides external access to power supply voltages, RS-232 data lines, remote displaydrive lines, remote zero lines, and remote print lines.
After installing PC board assembly P/N 27643-0030 as shown in assembly section, plug 3 pin cable assembly into P8. This enables user to have remote zero capabilities within the Model 715. Be sure to match dip switch selections between receiver card and transmitter. Place antenna through small hole in the lower stainless steel assembly in order to function properly.
Programming Appendix
A new feature makes the Model 715 more easily configurable with J-Star and Eaton transducers than before. The added J-Star/Eaton interface protocol slightly modifies display and keying sequences for accessing Configuration Mode, Calibration Mode, and Internal Raw Counts Mode. The Model 715 has added two more (716E, 717E) menus to make converting between different Weigh-Tronix applications even more flexible than it was with the 700E model.
Configuration for J-Star, Eaton and Two Additional
Weigh-Tronix Menus
New displays and keying sequences require modified instructions. Modified documentation presented in the PROGRAMMING APPENDIX applies whenever you change the transducer interface on your system, and the change means a different transducer trade name than that previously configured for. This includes switching from J-Star or Eaton transducers back to Weigh-Tronix Weigh Bars.
In applications where Weigh-Tronix Weigh Bars are retained, original documentation applies.
When switching the transducer interface on a system means a change in transducer trade names, follow the modified procedure shown below and illustrated in FIGURE 2 to access Configuration Mode, Calibration Mode, and Internal Raw Counts Mode.
Always execute the steps below before keying in a new Configuration Code or trying to calibrate a transducer with a different trade name.
Key in 715 and press E key for
two seconds. . . Currently configured menu is displayed: For Weigh-Tronix, 715E, 716E; for Eaton, EAtn 717E; for J- Star, JStr.
You now have two choices, A or B:
If displayed menu is correct, skip to step 4. OR
If displayed menu does not correspond with new transducer interface, press and hold E key until next menu appears.
If each of the Weigh-Tronix menus have been configured for different applications, it is possible to bounce between the 715E, 716E, and 717E menus. Press 715, 716, or 717 and hold the E key until the optional menu is displayed.
Continue keying as appropriate to access desired modes.
Configuration Code Charts
Using the Eaton Configuration Code Chart (Table 6)
Using the J-Star Configuration Code Chart (Table 7)
To determine an appropriate Configuration Code for J-Star or Eaton trans- ducers, consult the Configuration Code Charts in Tables 6 or 7.
The Eaton Configuration Code Chart (See Table 6) shows you how selec- tions listed below for five configurable parameters would be represented by a configuration code of 4151. Notice that each digit in the code represents a parameter, and the value of each digit represents a selection for that parameter as shown in Table 8: Eaton Configuration Code Chart.
Table 6 Eaton: A Sample Set of Configuration Choices and Its Corresponding Configuration Code, 4151 | |
Parameter Represented by | Selection Config. Code |
Sensitivity 2nd digit | .034/1K 4 |
Capacity x Increment Size 3rd digit | 20K x 2 1 |
Warning Alarm Factor 4th digit | 25 = 50lb / 2 5 |
Units of Measure | |
& Auto Hold Mode 5th digit | lb/ON 1 |
CODE 4151 uses 2nd, 3rd, 4th and 5th of five available digits. |
The J-Star configuration Code Chart (See Table 7) shows you how selec- tions listed below for five configurable parameters would be represented by a Configuration Code of 3151. Notice that each digit in the code represents a parameter, and the value of each digit represents a selection for that parameter as shown in Table 9: J-Star Configuration Code Chart.
Table 7 J-Star: A Sample Set of Configuration Choices and Its Corresponding Configuration Code, 3151 | |
Parameter Represented by | Selection Config. Code |
Axle Size 2nd digit | 2-1/8" 3 |
Capacity x Increment Size 3rd digit | 20K x 2 1 |
Warning Alarm Factor 4th digit | 25 = 50lb / 2 5 |
Units of Measure | |
& Auto Hold Mode 5th digit | lb/ON 1 |
CODE 3151 uses 2nd, 3rd, 4th and 5th of five available digits. |
Table 8: Eaton Configuration Code Chart
0: .068/100 (5-pin) | 2K x .1 | 2K x .2 | 2K x .5 | 20K x 1 | 20K x 2 | 20K x 5 |
1: .034/100 | 2K x .1 | 2K x .2 | 2K x .5 | 20K x 1 | 20K x 2 | 20K x 5 |
2: .017/100 | 2K x .1 | 2K x .2 | 2K x .5 | 20K x 1 | 20K x 2 | 20K x 5 |
3: .068/1K | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
4: .034/1K | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
5: .017/1K (5-pin) | 200K x 10 | 200K x 20 | 200K x 50 | 200K x 100 | 200K x 200 | 200K x 500 |
6: .100/1K (5-pin) | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
TRANSDUCER DESCRIPTION: SENSITIVITY
CAPACITY x INCREMENT SIZE
19: Custom Sensitivity and Capacity x Increment (See text: Configuring Custom Sensitivity with Capacity x Increment Size.)
WARNING ALARM FACTOR 0: 0
1: 2
2: 5
3: 10
4: 20
5: 25
6: 30
7: 40
8: 50
FIELD RECALIBRATION
To insure accuracy, recalibrate the scale system with a knownweightafterreconfiguringforEaton(SeeCalibratingUsingaKnownWeight, p. 10).
9: Custom (See text: Configuring Custom Warning Alarm Factor.)
0: lb No
1: lb Yes
2: kg No
3: kg Yes
_ 4 1 5 1 is the Configuration Code number representing selections cited in USING EATON TABLE 6.
1 2 3 4 5 are place numbers for each digit in Configuration Code.
Symbols Used in Table 6 K Multiply by 1000
@ Divide desired Warning Alarm Value by selected Increment Size to find Warning Alarm Factor.
Table 9: J-Star Configuration Code Chart
0: 1" | 2K x .1 | 2K x .2 | 2K x .5 | 20K x 1 | 20K x 2 | 20K x 5 |
1: 1" (Poly) | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
2: 1-7/8" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
2" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
3: 2-1/8" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
2-1/2" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
2-7/8" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
3-3/4" | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
4: CT | 20K x 1 | 20K x 2 | 20K x 5 | 200K x 10 | 200K x 20 | 200K x 50 |
TRANSDUCER DESCRIPTION: AXLE SIZE
CAPACITY x INCREMENT SIZE
19: Custom Sensitivity and Capacity x Increment (See text: Configuring Custom Sensitivity with Capacity x Increment Size.)
WARNING ALARM
FACTOR
0: 0
1: 2
2: 5
3: 10
4: 20
5: 25
6: 30
7: 40
8: 50
FIELD CALIBRATION
To insure accuracy, recalibrate the scale system with a known weight after reconfiguring for J-Star
(See Calibrating Using a Known Weight, p. 10).
9: Custom (See text: Configuring Custom Warning Alarm Factor.)
0: lb No
1: lb Yes
2: kg No
3: kg Yes
_ 3 1 5 1 is the Configuration Code number representing selections cited in USING TABLE 1: CONFIGURATION CODE EXAMPLE.
1 2 3 4 5 are place numbers for each digit in Configuration Code.
Symbols Used in Table 8 K Multiply by 1000
@ Divide desired Warning Alarm Value by selected Increment Size to find Warning Alarm Factor.
29MODEL 715 INDICATOR 12VDC
SYSTEM BLOCK DIAGRAM
MODEL 715 INDICATOR 12VDC
PARTS AND ASSEMBLY
ITEM NO. |
DESCRIPTION |
W-T P/N |
QTY |
1 | Mounting Bracket | 11899-0043 | 1 |
2 | Rubber Mount | 17807-0058 | 4 |
3 | Neoprene Washer | 27357-0010 | 4 |
4 | Enclosure | 47017-0010 | 1 |
5 | Main Pc Board | 47301-0015 | 1 |
6 | Main Pc Board W/ Rs-232, R/D | 47301-0023 | 1 |
7 | “Remote Zero” Receiver Kit (Optional) | 27643-0030 | 1 |
8 | Bezel | 21356-0014 | 1 |
9 | Screw, #10-32 x .75"L | 18087-0073 | 12 |
10 | Keypad Ass'y | 47697-0025 | 1 |
11 | Alarm Light Ass'y | 24129-0014 | 1 |
12 | Screw, #10-32 x .31"L | 14473-0348 | 8 |
13 | Plastic Cap (Threaded) | 15351-0029 | 1 |
14 | Xm710, Hand-Held Remote | 27809-0014 | 1 |
15 | Pnl Assy, Single W/B Conn (7-Pin) | 52956-0054 | 1 |
Pnl Assy, Single W/B Conn (7-Pin) W/Rd or RS232 | 52956-0104 | 1 | |
Pnl Assy 3-W/B Conn (4-Pin) | 52956-0047 | 1 | |
Pnl Assy 3-W/B Conn (4-Pin) W/Rd or RS232 | 52956-0096 | 1 | |
Pnl Assy 3-W/B Conn (5-Pin) | 52956-0039 | 1 | |
Pnl Assy 3-W/B Conn (5-Pin) W/Rd or RS232 | 52956-0088 | 1 | |
Pnl Assy 4-W/B Conn (4-Pin) | 52956-0021 | 1 | |
Pnl Assy 4-W/B Conn (5-Pin) | 52956-0013 | 1 | |
Pnl Assy 4-W/B Conn (4-Pin) W/Rd or RS232 | 52956-0070 | 1 | |
Pnl Assy 4-W/B Conn (5-Pin) W/Rd or RS232 | 52956-0062 | 1 | |
16 | Fuse Holder | 15455-0016 | 2 |
17 | Fuse (5 AMP) | 15453-0042 | 2 |
MODEL 715 INDICATOR 12VDC WEIGH BAR PANEL ASSEMBLY AND CABLE / CONNECTOR SCHEMATICS
MODEL 715 INDICATOR 12VDC
KEYPAD OVERLAY P/N 47697-0025 AND SCHEMATIC
EXTERNAL INTERFACE CABLE PIN-OUTS
MODEL 715 INDICATOR 12VDC
MAIN PC BOARD A/D ASSEMBLY
P/N 47301 –0015, -0023 & COMPONENTS
SYMBOL | DESCRIPTION | W-T P/N |
R20,21 | RESISTOR, 182 OHM | 15677-2188 |
R22 | RESISTOR, 470 K | 14477-1375 |
R23 | RESISTOR, 510 K | 14477-1383 |
R24, 25 | RESISTOR, 3.3 K | 14477-0856 |
R26,30 | RESISTOR, 2.0 K | 14477-0807 |
R27 | RESISTOR, 390 OHM | 15670-0635 |
R28 | RESISTOR, 750 OHM | 15670-0700 |
R29 | RESISTOR, 910 OHM | 14477-0724 |
R31-34 | RESISTOR, 15 K | 14477-1011 |
R35 | RESISTOR, 0.475 OHM | 46602-0179 |
R41,42 | RESISTOR, 10.0 K | 15677-3855 |
R44 | RESISTOR, 3.9 K | 14477-0872 |
R45,51 | RESISTOR, 4.7 OHM | 14477-0179 |
R46,47 | RESISTOR, 22 OHM | 14477-0336 |
R48 | RESISTOR, 78.7 K | 15677-4713 |
R49 | RESISTOR, 13.0 K | 15677-3962 |
R50 | -------------- NOT USED ---------- | --------------- |
R52 | RESISTOR, 10 MEG | 14477-1698 |
R53 | RESISTOR, 5.1 K | 14477-0906 |
R54-57 | RESISTOR, 120 OHM | 14477-0518 |
R58-62 | RESISTOR, 22K | 14477-1052 |
R63,64 | RESISTOR, 39 K | 14477-1110 |
R65 | RESISTOR, 22 MEG | 14477-1771 |
R66 | RESISTOR, 5.23 K | 15673-3586 |
R67 | RESISTOR, 1.0 K | 15673-2893 |
RN1 | RESISTOR NETWORK, 5.62/301K | 23256-0011 |
RN2-4 | RESISTOR NETWORK, 9 X 10 K | 17852-0045 |
RN5,6 | RESISTOR NETWORK, 4 X .1 K | 17852-0037 |
RV1 | VARISTOR, V827A12 | 16046-0028 |
U1 | IC, 7652 | 14323-0464 |
U2,3,5,7 | IC, 308A | 14323-0266 |
U4 | IC, DG200 | 14323-0241 |
U6 | IC, 311 | 14323-0258 |
U8 | VOLTAGE REGULATOR 7660 | 15658-0227 |
U9 | IC, 690 | 15657-0673 |
U10 | VOLTAGE REGULATOR 7805 | 15658-0011 |
U11 | IC, MC68HC11A1P | 15657-0632 |
U12 | IC, 74HC573 | 18080-0559 |
U13 | IC, PROGRAMMED MOD 715 | 27657-0058 |
U14 | IC, 74HC00 | 18080-0013 |
U15 | IC, 5453N | 15657-0624 |
U16 | IC, 74HC14 | 18080-0096 |
U17 | IC, MAX232CPE | 14323-0621 |
U19 | REGULATOR, 2941T | 15658-0359 |
VR2 | DIODE, ZENER 1N5353B 16V | 15669-0158 |
XDS1 | SOCKET, SIP 25 PIN | 17847-1017 |
XDS1 | SOCKET, SIP 10 PIN | 17847-1066 |
XU13 | SOCKET, IC 28 PIN | 14361-0079 |
XU18 | HEAT SINK | 15556-0014 |
Y1 | CRYSTAL, 8.00 MHZ | 16125-0212 |
SYMBOL | DESCRIPTION | W-T P/N |
B1 | BATTERY, 3.6 V | 23957-0021 |
BL1 | BACKLITE, DISPLAY | 27459-0041 |
C1,2,5,6,9, 11,12,14,15 ,22,23,26,2 9,30,32,33, 47,48,50,53 ,59,60, 61 |
CAP. 0.1 UF 100 V |
15623-0120 |
C3,4 | CAP. .22 UF 50 V | 23267-0117 |
C7,8,18,19, 27,28 |
CAP. 1.0 UF 35 V |
22327-2519 |
C10,38-46, 49,55-58 |
CAP. 47 UF 16 V |
17993-0094 |
C13,17,21, 37 |
CAP. 100PF 200 V |
15619-0134 |
C16 | CAP. 1.0 UF 50 V | 23267-0158 |
C20 | CAP. .15 UF 100 V | 18083-0127 |
C24 | CAP. 4700PF 100 V | 15620-0018 |
C25 | CAP. 0.01 UF 100 V | 15620-0123 |
C31 | CAP. 470 UF 35 V | 17995-0134 |
C34 | CAP. 1000 UF 35 V | 17995-0142 |
C35 | CAP. 10 UF 35 V | 22327-2618 |
C36 | CAP. 0.01 UF 100 V | 15620-0123 |
C51,52 | CAP. 22PF 200 V | 15619-0050 |
C54 | CAP. 1000 PF 200 V | 15619-0258 |
CR1,4 | DIODE, 1N4142 | 15668-0076 |
CR3 | DIODE, 1N5817 | 15668-0324 |
CR2,6,7 | DIODE, 1N4004 | 15668-0043 |
CR5 | DIODE, 1N4148 | 15668-0035 |
CR8 | -------------- NOT USED---------- | --------------- |
DS1 | DISPLAY, LCD | 22329-0040 |
DS2,3 | LED, H-2000 | 27472-0010 |
FL1-6 | FILTER, EMI | 46547-0011 |
K1 | RELAY | 17982-0022 |
L1,2 | CHOKE, 10 UHY | 15779-0015 |
P1,2 | CONN. 6 PIN | 17794-0053 |
P3 | CONN. 2 PIN | 17794-0012 |
P4 | CONN. 12 PIN | 17731-0117 |
P5 | CONN. 14 PIN | 17794-0137 |
P6,7 | CONN. 2 PIN | 17734-0015 |
P8 | CONN. 3 PIN | 17794-0020 |
Q1 | TRANSISTOR, VN10KM | 15665-0152 |
Q2 | TRANSISTOR, TIP122 | 16271-0032 |
Q3-5 | TRANSISTOR, 2N3904 | 15665-0012 |
R5-7,12,17, 19,43 |
RESISTOR, 100 OHM |
14477-0492 |
R8 | RESISTOR, 9.09 K | 17873-1899 |
R9 | RESISTOR, 64.9 K | 17873-2715 |
R10 | RESISTOR, 196 K | 17873-3176 |
R11 | RESISTOR, 8.06 K | 15677-3764 |
R13,14 | RESISTOR, 37.4 K | 17873-2483 |
R15,16 | RESISTOR, 42.2 K | 17873-2533 |
R18 | RESISTOR, 110 K | 15673-4857 |
715MB
MODEL 715 INDICATOR 12VDC
MAIN PC BOARD A/D SCHEMATIC0
RD712 / 712XL REMOTE DISPLAY
ASSEMBLY P/N 48406 AND PARTS LIST
ITEM NO. |
DESCRIPTION |
W-T P/N |
QTY |
1 | Mounting Bracket | 11899-0043 | 1 |
2 | Rubber Mount | 17807-0058 | 4 |
3 | Neoprene Washer | 27357-0012 | 4 |
4 | Enclosure | 47017-0010 | 1 |
5 | Main Pc Board Assy W/Display (RD712) | 48445-0010 | 1 |
6 | Main Pc Board Assy W/Display (RD712XL) | 48865-0029 | 1 |
7 | “Remote Zero” Receiver Assy (Optional) | 51763-0018 | 1 |
8 | Screw, #6-32 x .25"L | 14473-0223 | 4 |
9 | Lock Washer #6 | 14474-0032 | 4 |
10 | Rd712 Keypad/Backer Plate Assy (incl.: keypad, backer | ||
plate, gasket) | 48279-0029 | 1 | |
11 | Rd712xl Keypad/Backer Plate Assy (incl.: keypad, backer plate, gasket) |
48715-0021 |
1 |
12 | Cable Assy 15' | 48412-0027 | 1 |
Cable Assy 30' | 48412-0019 | 1 | |
13 | Bezel | 21356-0014 | 1 |
14 | Screw #10-32 x .75"L | 18087-0073 | 12 |
15 | Xm710-L Hand Held Remote (optional) | 51760-0011 | 1 |
RD712 REMOTE DISPLAY
SYMBOL | DESCRIPTION | W-T P/N |
BL1 | BASCKLIGHT DSPL FIB OP | 27459-0058 |
C1, 2 | CAP .1 UF 100 V | 15623-0120 |
C3 | CAP 100 UF 25 V | 17994-0101 |
C4 | CAP 47 UF 10 V | 17992-0095 |
C5, 6 | CAP 0.1 UF 50 V | 46684-0048 |
C7 | CAP 1000 PF 100 V | 46684-0022 |
CR1 | DIO 1N4142 | 15668-0076 |
DS1 | DISPLAY LCD 2" | 46671-0019 |
DS2,3 | LED | 27472-0010 |
L1 | CHOKE 10 UHY | 15779-0015 |
P1 | CONN 10 PIN | 17794-0095 |
P2 | CONN 12 PIN | 17731-0117 |
P3 | CONN 3 PIN | 17794-0020 |
Q1 | TRANS TIP122 | 16271-0032 |
SYMBOL | DESCRIPTION | W-T P/N |
R1 | RES 303 K | 14477-0856 |
R2 | RES 1.0 K | 14477-0732 |
R3 | RES 510 K | 14477-1383 |
R4,6,10,14 | RES 10 K | 14477-0971 |
R5, 7 | RES 22 K | 14477-1052 |
R8,9,12,13 | RES 120 OHM | 14477-0518 |
R11 | RES 330 OHM | 15670-0619 |
RV1 | VARIS V82ZA12 | 16046-0028 |
U1 | IC 74HC14 | 18080-0096 |
U2 | IC 5453N | 15657-0624 |
XDS1 | SOCKET 25 PIN | 17847-1017 |
XDS1 | SOCKET 10 PIN | 17847-1066 |
DISPLAY PC BOARD A/D P/N 48445-0010, COMPONENTS AND SCHEMATIC
SYMBOL | DESCRIPTION | W-T P/N |
BL1 | BACK LITE DSPL FIB OP | 27459-0066 |
C1,2 | CAP .1 UF 100 V | 15623-0120 |
C3 | CAP 100 UF 25 V | 17994-0101 |
C4 | CAP 47 UF 16 V | 17993-0094 |
C5 | CAP 47 UF 10 V | 17992-0095 |
C6,7 | CAP 0.1 UF 50 V | 46684-0048 |
C8 | CAP 1000 PF 100 V | 46684-0022 |
CR1 | DIO 1N4142 | 15668-0076 |
DS1 | DISPLAY LCD 2" | 46643-0014 |
DS2,3,4 | LED | 27472-0010 |
L1 | INDUCTOR 10 UHY | 15779-0015 |
P1 | CONN 10 PIN | 17794-0095 |
P2 | CONN 12 PIN | 17734-0114 |
P3 | CONN 3 PIN | 17794-0020 |
SYMBOL | DESCRIPTION | W-T P/N |
Q1 | TRANS TIP122 | 16271-0032 |
R1 | RES 303 K | 14477-0856 |
R2 | RES 1.0 K | 14477-0732 |
R3 | RES 1 MEG | 14477-1458 |
R4,6,10,14 | RES 10 K | 14477-0971 |
R5,7 | RES 22 K | 14477-1052 |
R8,9,12,13 | RES 120 OHM | 14477-0518 |
R11 | RES 33 OHM | 14477-0377 |
U1 | IC 74HC14 | 18080-0096 |
U2 | IC 5453N | 15657-0624 |
U3 | VOLT REG | 15658-0292 |
XDS1 | SOCKET 25 PIN | 17847-1017 |
XDS1 | SOCKET 10 PIN | 17847-1066 |
RD712XL REMOTE DISPLAY DISPLAY PC BOARD P/N 48865-0029, COMPONENTS AND SCHEMATIC
XM710-L WIRELESS REMOTE INSTALLATION
The XM710-L hand-held transmitter provides the ability to remotely perform a zero/reset command and can be installed
in MODEL 715 or RD712 / 712XL. The XM710-L wireless remote system is similar to those used on garage door openers. For selectable security, the transmitter and receiver assembly boards have programmable security code switches.
The eight position switch is used to determine the security code. These can be programmed to a different code, but the programmed code must match on both the transmitter and receiver. The unit is shipped from the factory with the following settings:
1, 3, 5, 7 = ON (closed)
2, 4, 6, 8 = OFF (open)