Калибровка весов типа BW (CAS)

Калибровка весов типа BW (CAS)

SPAN CALIBRATION FOR BW-SERIES

Make sure scale is OFF. Open the head assembly via the RELEASE LATCH located under- neath the head assembly opposite the KEYPAD.

Remove the CALIBRATION BOLTS and the calibration plate located on the metal frame assembly to the left of the RELEASE LATCH.

You will see a push button MICRO-SWITCH. Press and hold down the MICRO SWITCH as you press the ON/OFF key.

The WEIGHT display should read “SET” for a second and then it will continuously scroll a message.

 

Press the TARE key. The display will flash CAL 1 and then show the MAX Capacity of the scale in lbs. To change the value for the CAL settings first press the ZERO/MODE key to clear the number. Press the NET/GROSS key to increment the first number. Press the TARE key to multiply the number by 10. When finished press the kg/lb key.

 

The display will flash CAL 2 and then show the MAX Capacity in KG. If the setting is incorrect then go to step 6. Press the kg/lb key.

 

The display will flash CAL 3 and then show the MINIMUM Capacity in lbs. if the setting is incorrect then press the NET/GROSS key to go through the preset minimum values. Be sure to select your scales minimum value. A smaller value will not be accepted. When finished press the kg/lb key.

 

The display will flash CAL 4 and then show the unit of weight to be used to calibrate the scale. To change this setting press the NET/GROSS key to select KG or LBS. when finished press the kg/lb key.

 

The display will flash CAL 5 and then show the amount of weight to calibrate with. (the defalt value is max capacity) To change this setting see step 6. At least 10% of max capacity must be used to calibrate. When finished press the kg/lb key.

 

The display will flash CAL 6 and then show Uload. Be sure there is nothing on the platter and press the kg/lb

key.

 

After a few seconds the display will flash CAL 7 and show load. Put on the platter the amount of weight you entered in step 10 and press the kg/lb key.

After a few seconds the display will flash END an the display SET and go back to the scrolling message. At this point turn scale off and back on again. Test the accuracy of the scale.

 

If scale is NOT 100% accurate, unplug scale and repeat steps 1 to 11.

 

If scale IS 100% accurate: You are done.

 

CAL MODE PROGRAMMING FOR BW-SERIES

 

Make sure scale is OFF. Open the head assembly via the RELEASE LATCH located under- neath the head assembly opposite the KEYPAD.

 

Remove the CALIBRATION BOLTS and the calibration plate located on the metal frame assembly to the left of the RELEASE LATCH.

 

You will see a push button MICRO-SWITCH. Press and hold down the MICRO SWITCH as you press the ON/OFF key.

The WEIGHT display should read “SET” for a second and then it will continuously scroll a message.

 

Press the TARE key and then press the KG/LB key until the display reads “ULoAd”.

Make sure that the platter is properly placed on the scale platform.

 

Press the KG/LB key. The scale will show some dashes and then show “LoAd”.

Once the displays read “LoAd” put the FULL CAPACITY OF WEIGHT 30, 60, 150, or 300 lb (15, 30, 60, or 150 kg) on the platter and press the KG/LB key.

Once the scale displays “End”, remove all weights from the platter.

 

Press the ON/OFF key. Once the scale is OFF press the ON/OFF key again.

The scale will do an LCD segment test and go to ZERO.

 

Test the accuracy of the scale.

 

If scale is NOT 100% accurate, unplug scale and repeat steps 1 to 11.

 

If scale IS 100% accurate: unplug scale and you are done.